4406 ISMRM Abstract

Design of a universal RF-shimming drive mode for head and neck imaging at 7 Tesla using a 16-channel pTx array

Accepted
Ehsan Kazemivalipour 1,2, Markus W May3,4, D. Rangaprakash1,2, Berkin Bilgic1,2,5, Jason Stockmann1,2, Robert L Barry1,2, Boris Keil6,7, Lawrence L Wald1,2,5, Bastien Guerin1,2
1A. A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Boston, United States of America
2Harvard Medical School, Boston, United States of America
3Erwin L. Hahn Institute for MRI, University Duisburg-Essen, Essen, Germany
4High-Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany
5Harvard-MIT Division of Health Sciences and Technology, Cambridge, United States of America
6Institute of Medical Physics and Radiation Protection, Department of Life Science Engineering, Mittelhessen University of Applied Sciences, Gießen, Germany
7Center for Mind, Brain and Behavior (CMBB), Philipps-University Marburg, Marburg, Germany
Presenting Author: Ehsan Kazemivalipour
Full abstract & presentation

The full text, figures, and any recorded presentation for this abstract are not shown here.

This content is freely available and no longer restricted to registered attendees. Log in or create a free account to view it.

Log in

References

1. May MW, Hansen SJD, Mahmutovic M, et al. A patient-friendly 16-channel transmit/64-channel receive coil array for combined head-neck MRI at 7 Tesla. Magn Reson Med. 2022;88(3):1419-1433.
2. Massire A, Cloos MA, Luong M, et al. Thermal simulations in the human head for high field MRI using parallel transmission. J Magn Reson Imaging. 2012;35(6):1312-1321.
3. Makris N, Angelone L, Tulloch S, et al. MRI-based anatomical model of the human head for specific absorption rate mapping. Med Biol Eng Comput. 2008;46(12):1239-1251.
4. Noetscher GM. The CAD-Compatible VHP-Male Computational Phantom. In Makarov SN, Noetscher GM, Nummenmaa A, (Eds). Brain and Human Body Modeling 2020: Computational Human Models Presented at EMBC 2019 and the BRAIN Initiative(R) 2019 Meeting. Cham (CH) 2021:309-323.
5. Makarov SN, Noetscher GM, Yanamadala J, et al. Virtual Human Models for Electromagnetic Studies and Their Applications. IEEE Rev Biomed Eng. 2017;10:95-121.
6. Guerin B, Gebhardt M, Cauley S, et al. Local specific absorption rate (SAR), global SAR, transmitter power, and excitation accuracy trade-offs in low flip-angle parallel transmit pulse design. Magn Reson Med. 2014;71(4):1446-1457.
7. Eichfelder G, Gebhardt M. Local specific absorption rate control for parallel transmission by virtual observation points. Magn Reson Med. 2011;66(5):1468-1476.
8. Lee J, Gebhardt M, Wald LL, et al. Local SAR in parallel transmission pulse design. Magn Reson Med. 2012;67(6):1566-1578.
9. Grissom W, Yip CY, Zhang Z, et al. Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magn Reson Med. 2006;56(3):620-629.
10. Brunner DO, Pruessmann KP. Optimal design of multiple-channel RF pulses under strict power and SAR constraints. Magn Reson Med. 2010;63(5):1280-1291.
11. Gras V, Vignaud A, Amadon A, et al. Universal pulses: A new concept for calibration-free parallel transmission. Magn Reson Med. 2017;77(2):635-643.

Cite this abstract